Brookfield Engineering Laboratories, Inc. Page 1 Manual No. M/92-021-O604BROOKFIELD DIGITAL VISCOMETERMODEL DV-I+Operating InstructionsManual No.
Brookfield Engineering Laboratories, Inc. Page 10 Manual No. M/92-021-O604cP 123E3 70.1°F0.0RPM % 0.0cP 0.0 S010.0RPM % 0.0(with Temperat
Brookfield Engineering Laboratories, Inc. Page 11 Manual No. M/92-021-O604NOTE: You have approximately three seconds in which to press the SELECT
Brookfield Engineering Laboratories, Inc. Page 12 Manual No. M/92-021-O604 II.3 Speed Selection & SettingTable II-1 shows the available speed
Brookfield Engineering Laboratories, Inc. Page 13 Manual No. M/92-021-O604DV-I+ is currently running. The user may operate at a given speed and p
Brookfield Engineering Laboratories, Inc. Page 14 Manual No. M/92-021-O604CGS SIViscosity cP mPa.s II.6 Temperature Display in °F or °C Selection
Brookfield Engineering Laboratories, Inc. Page 15 Manual No. M/92-021-O604 II.8 OperationThe following procedure is outlined for making a viscosi
Brookfield Engineering Laboratories, Inc. Page 16 Manual No. M/92-021-O6043. Next, the user selects the spindle number corresponding to the spindl
Brookfield Engineering Laboratories, Inc. Page 17 Manual No. M/92-021-O6043. At this point, the user need only press the MOTOR ON/OFF key to begin
Brookfield Engineering Laboratories, Inc. Page 18 Manual No. M/92-021-O604 TIMED TORQUESET TORQUE: 00%Figure II-22Using the UP and DOWN ARROW key
Brookfield Engineering Laboratories, Inc. Page 19 Manual No. M/92-021-O604CP 123456789TIMED TORQ DONEFigure II-26The display will switch between t
Brookfield Engineering Laboratories, Inc. Page 2 Manual No. M/92-021-O604TABLE OF CONTENTS I. INTRODUCTION ...
Brookfield Engineering Laboratories, Inc. Page 20 Manual No. M/92-021-O604Appendix A - Cone/Plate Viscometer Set-UpThis Cone/Plate version of the
Brookfield Engineering Laboratories, Inc. Page 21 Manual No. M/92-021-O604Bath/CirculatorBathInletBathOutletSampleCup(CPE-44YorCPE-44P)CupOutletCu
Brookfield Engineering Laboratories, Inc. Page 22 Manual No. M/92-021-O604 A.3 SETTING THE GAP1. Move the toggle switch to the right; this willtur
Brookfield Engineering Laboratories, Inc. Page 23 Manual No. M/92-021-O604Cone Sample VolumePart No.CPE-40 0.5 mlCPE-41 2.0 mlCPE-42 1.0 mlCPE-51
Brookfield Engineering Laboratories, Inc. Page 24 Manual No. M/92-021-O604 Appendix B - Viscosity Ranges LV(#1-4) and RV,HA,HB(#1-7) ViscometersVi
Brookfield Engineering Laboratories, Inc. Page 25 Manual No. M/92-021-O604 Small Sample Adapter and ThermoselSSA and Viscosity (cP) ShearThermosel
Brookfield Engineering Laboratories, Inc. Page 26 Manual No. M/92-021-O604 Spiral AdapterShearRateSpiral Spindle LVDV-I+ RVDV-I+ HADV-I+ HBDV-I+ s
Brookfield Engineering Laboratories, Inc. Page 27 Manual No. M/92-021-O604all particle movement is in layers directed by the shearing force. For
Brookfield Engineering Laboratories, Inc. Page 28 Manual No. M/92-021-O604 Appendix C - Variables in Viscosity MeasurementAs with any instrument m
Brookfield Engineering Laboratories, Inc. Page 29 Manual No. M/92-021-O604 Viscometer Related VariablesMost fluid viscosities are found to be non-
Brookfield Engineering Laboratories, Inc. Page 3 Manual No. M/92-021-O604 I. INTRODUCTIONThe Brookfield DV-I+ Viscometer measures fluid viscosity
Brookfield Engineering Laboratories, Inc. Page 30 Manual No. M/92-021-O604 Appendix D - Spindle and Model CodesEach spindle has a two digit code w
Brookfield Engineering Laboratories, Inc. Page 31 Manual No. M/92-021-O604Table D-2 lists the model codes and spring torque constants for each vis
Brookfield Engineering Laboratories, Inc. Page 32 Manual No. M/92-021-O604 Appendix E - Calibration ProceduresThe accuracy of the DV-I+ is verifi
Brookfield Engineering Laboratories, Inc. Page 33 Manual No. M/92-021-O604 Brookfield Viscosity Standard Fluid - General InformationWe recommend t
Brookfield Engineering Laboratories, Inc. Page 34 Manual No. M/92-021-O604 Calibration Procedure for a Small Sample AdapterWhen a Small Sample Ada
Brookfield Engineering Laboratories, Inc. Page 35 Manual No. M/92-021-O604 Calibration Procedure for UL AdapterWhen a UL Adapter is used, the wate
Brookfield Engineering Laboratories, Inc. Page 36 Manual No. M/92-021-O604 Calibration Procedure for a Helipath Stand and T-Bar SpindlesT-Bar spin
Brookfield Engineering Laboratories, Inc. Page 37 Manual No. M/92-021-O6044) With the viscometer stopped, remove the sample cup and place the visc
Brookfield Engineering Laboratories, Inc. Page 38 Manual No. M/92-021-O604The current guard leg is a band of metal in the shape of the letter U wi
Brookfield Engineering Laboratories, Inc. Page 39 Manual No. M/92-021-O604 Appendix F - Special Speed SetsThe following special speeds sets are av
Brookfield Engineering Laboratories, Inc. Page 4 Manual No. M/92-021-O604 I.1 ComponentsComponent Part Number QuantityDV-I+ Viscometer varies 1Mod
Brookfield Engineering Laboratories, Inc. Page 40 Manual No. M/92-021-O604 Appendix G - Communications234567891No ConnectionNo ConnectionAnalog G
Brookfield Engineering Laboratories, Inc. Page 41 Manual No. M/92-021-O604Figure H-1*Provided with instruments shipped after October 1, 1999. Rep
Brookfield Engineering Laboratories, Inc. Page 42 Manual No. M/92-021-O604 UnpackingCheck carefully to see that all the components are received wi
Brookfield Engineering Laboratories, Inc. Page 43 Manual No. M/92-021-O604Appendix I - DVE-50A Probe ClipTpProbe Clip DVE-50A is supplied with the
Brookfield Engineering Laboratories, Inc. Page 44 Manual No. M/92-021-O604 Appendix J - Fault Diagnosis and Troubleshooting Spindle Does Not Rotat
Brookfield Engineering Laboratories, Inc. Page 45 Manual No. M/92-021-O604 Inaccurate Readings❏ Verify Spindle, Speed and Model selection❏ Verify
Brookfield Engineering Laboratories, Inc. Page 46 Manual No. M/92-021-O604 Appendix K - Warranty Repair and Service WarrantyBrookfield Viscometers
Brookfield Engineering Laboratories, Inc. Page 47 Manual No. M/92-021-O604❏ Remove and return all spindles (properly packedfor shipping).❏ Clean e
Brookfield Engineering Laboratories, Inc. Page 48 Manual No. M/92-021-O604Providing us with the following information will help us to service your
Brookfield Engineering Laboratories, Inc. Page 49 Manual No. M/92-021-O604Appendix L - Viscosity Test ReportVISCOSITY TEST REPORTSAMPLE MODEL SPIN
Brookfield Engineering Laboratories, Inc. Page 5 Manual No. M/92-021-O604 I.3 SpecificationsSpeeds: LVDV-I+: 0.0, 0.3, 0.6, 1.5, 3, 6, 12, 30, 60,
Brookfield Engineering Laboratories, Inc. Page 6 Manual No. M/92-021-O604Tp 4) Connect the optional RTD temperature probe to the temperature port
Brookfield Engineering Laboratories, Inc. Page 7 Manual No. M/92-021-O604SETSPEEDSELECTSPINDLESELECTSPINDLEMOTORON/OFFAUTORANGEFigure I-1 I.6 Key
Brookfield Engineering Laboratories, Inc. Page 8 Manual No. M/92-021-O604 I.7 CleaningBe sure to remove spindle from instrument prior to cleaning.
Brookfield Engineering Laboratories, Inc. Page 9 Manual No. M/92-021-O604 II. GETTING STARTED II.1 Auto ZeroBefore readings may be taken, the Visc