Brookfield Engineering Labs., Inc. Page 69 Manual No. M/97-164-F1 102
Inaccurate Readings
Verify spindle, speed and model selection.
If % readings are under-range (less than 10%), the display will flash; change spindle and/or
speed.
EEE on the digital display means the unit is over-range (greater than 100%); reduce speed
and/or change spindle.
Verify test parameters: temperature, container, volume, method. Refer to:
•“More Solutions to Sticky Problems, Section III
•“DV-II+ Programmable Viscometer Operating Instructions, Appendix C, Variables in
Viscosity Measurements.
Perform a calibration check; follow the instructions in Appendix E.
Verify tolerances are calculated correctly.
Verify the calibration check procedures were followed exactly.
If the unit is found to be out of tolerance, the unit may be in need of service. See Appendix K for
details on How to Return Your Viscometer.
Viscometer Will Not Return to Zero
Viscometer is not level
Check with spindle out of the sample
Adjust the laboratory stand
Pivot point or jewel bearing faulty
Perform calibration check
Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair
Display Reading Will Not Stabilize
Special characteristic of sample fluid. There is no problem with the viscometer.
Refer to Appendix C
Check for erratic spindle rotation.
Verify power supply
Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair
Bent spindle or spindle coupling.
Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair
Temperature fluctuation in sample fluid
Use temperature bath control.