Brookeld Engineering Labs., Inc. Page 34 Manual No. M13-2100
atStepEnd.Duringthetest,thetotaltimerequiredtoreach200cPis65seconds.
Total pointscollectedwillbe7withthelastdatapointtaken at65seconds,an
averageofthe5secondsafterpoint#6.
Example8:EndConditionissettoViscosity=200cP,MultiPointDataIntervalis1minute.
AveragingDurationis20seconds.CheckthecheckboxtoAlsoCollectSingle
PointatStepEnd.Duringthetest,thetotaltimerequiredtoreach200cPis10
minutes40seconds.Totalpointscollectedwillbe11withthelastdatapointtaken
at10minutes40seconds,anaverageofthelast20secondsofthestep.
No Data: End Condition is met and no data is collected.
III.8 End Condition
ThecompletionofatestisdenedbytheEndCondition.EachtimethatyouenterCongure
ViscosityTest,theEndConditionwillbesettothelastvalueused.TheEndConditionparameter
orvaluescanbechangedbypressingtheEndConditionbutton.
WithintheEndConditionscreen,thecurrentlyselectedEndConditionparameterandvaluesare
displayed.TheEndConditioncanbechangedbypressingthebluedownarrowwithintheparameter
eld.SixEndConditionsareavailable:Time,#ofPoints,#ofRevolutions,Torque,Viscosity,
Temperature(seeFigureIII-10).
Figure III-10
Time:Thetestwillcompletewhenthespeciedamountoftimehaselapsed.Timeisentered
inHours,Minutes,andSeconds.
Hours: 0-99
Minutes: 0-59
Seconds: 0-59
Atimeofzerohours,minutesandsecondsmaybeselected.WiththisEndCondition,
theDV3TwilloperateattheselectedspeeduntiltheoperatorselectsStopTest.Data
willbecollectedaccordingtotheDataCollectionsetting(seeSectionIII.7).