Brookfield DV3T Rheometer Uživatelský manuál Strana 45

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 125
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 44
Brookeld Engineering Labs., Inc. Page 45 Manual No. M13-2100
Multi Point Averagingrequiresadenitionof1)DataInterval,thefrequencyofdatacollection
and2)AveragingDuration,theamountoftimeforwhichreadingswillbeaveraged.Thesetwo
parameterswillworkinconjunctiontogeneratemultipledatapointseachofwhichrepresent
theaverageofalldatameasuredduringtheaveragedurationwithinthespeciedDataInterval.
PostTestAveragingofdataoccursafterthetestiscomplete,throughtheResultsscreen.Averages
and standard deviation values can be generated for data collected in a single step or across several
steps.IfthetestusedtocollectdatautilizedLiveAveraging(describedabove),thenthePostTest
Averaging will produce an average value of averaged data points.
PostTestAveragingcanbespeciedintheReportsectionofCongureViscosityTest(seeSection
III.9).Whenconguredaspartofthetest,theaveragevalueswillbedisplayedimmediately
upontestconclusion.IfnotspeciedinReport,PostTestAveragingcanbeselectedfromthe
DownArrowintheResultsscreen(seeSectionIII.11).
PostTestAveragingofferstwooptions:StepAveragingandTestAveraging(seeFigureIII-20).
Figure III-20
Zobrazit stránku 44