Table III-1 Full Scale Range Viscosity of LOW TORQUE CAP Viscometer
Cone
Number
Cone Constant
Range
Shear Rate
Constant
FSR at
100 RPM
FSR at
500 RPM
FSR at
any RPM
01 1875 13.33N 0.83 0.17 1875/(22.7*N)
02 3750 13.33N 1.65 0.33 3750/(22.7*N)
03 7500 13.33N 3.30 0.66 7500/(22.7*N)
04 15000 3.33N 6.61 1.32 15000/(22.7*N)
05 30000 3.33N 13.22 2.64 30000/(22.7*N)
06 75000 3.33N 13.04 6.61 75000/(22.7*N)
07 3150 2N 1.39 0.28 3150/(22.7*N)
08 12500 2N 5.51 1.10 12500/(22.7*N)
09 50000 2N 22.03 4.41 50000/(22.7*N)
10 5000 5N 2.20 0.44 5000/(22.7*N)
Table III-2 Full Scale Range Viscosity of HIGH TORQUE CAP Viscometer
Cone
Number
Cone
Constant
Range
Shear
Rate
Constant
FSR
at
100 RPM
FSR
at
400 RPM
FSR
at
750 RPM
FSR
at
900 RPM
FSR
at
any RPM
01 1875 13.33N 18.75 4.69 2.50 2.08 1875/(N)
02 3750 13.33N 37.50 9.38 5.00 4.17 3750/(N)
03 7500 13.33N 75.00 18.75 10.00 8.33 7500/(N)
04 15000 3.33N 150.00 37.50 20.00 16.67 15000/(N)
05 30000 3.33N 300.00 75.00 40.00 33.33 30000/(N)
06 75000 3.33N 750.00 187.50 100.00 83.33 75000/(N)
07 3150 2N 31.50 7.88 N/A N/A 3150/(N)
08 12500 2N 125.00 31.25 N/A N/A 12500/(N)
09 50000 2N 500.00 125.00 N/A N/A 50000/(N)
10 5000 5N 50.00 12.50 6.67 5.56 5000/(N)
(Continued)