Brookfield Engineering Labs., Inc. Page 1 Manual No. M/94-310F
BROOKFIELD CAP VISCOMETERS
MODELS CAP 1000 and CAP 2000
Operating Instructions, version 1.20
Manual No. M/94-310-F699
SPECIALISTS IN THE
MEASUREMENT AND
CONTROL OF VISCOSITY
T
EL
508-946-6200
F
AX
508-946-6262
or 800-628-8139 (USA excluding MA)
I
NTERNET
http://www.brookfieldengineering.com
BROOKFIELD ENGINEERING LABORATORIES, INC.
11 Commerce Boulevard, Middleboro, MA 02346 USA
Zobrazit stránku 0
1 2 3 4 5 6 ... 36 37

Shrnutí obsahu

Strany 1 - Manual No. M/94-310-F699

Brookfield Engineering Labs., Inc. Page 1 Manual No. M/94-310FBROOKFIELD CAP VISCOMETERSMODELS CAP 1000 and CAP 2000Operating Instructions, version 1.

Strany 2 - TABLE OF CONTENTS

Brookfield Engineering Labs., Inc. Page 10 Manual No. M/94-310Fagain and the full scale viscosity is shown (Figure 9). Press RPM once more and the di

Strany 3 - I. INTRODUCTION

Brookfield Engineering Labs., Inc. Page 11 Manual No. M/94-310FNote: Entering 00 sets the manual run time mode. Pressing and holding the RUN key rota

Strany 4 - °C (4°F) to 40°C (149°F)

Brookfield Engineering Labs., Inc. Page 12 Manual No. M/94-310F5. Pressing and holding the STOP and RPM keys simultaneously executes the sequenceShear

Strany 5 - I.5 Key Functions

Brookfield Engineering Labs., Inc. Page 13 Manual No. M/94-310F III.1 Accuracy of MeasurementTable 3.1 indicates the accuracy of the viscosity measure

Strany 6 - °C=degrees centigrade

Brookfield Engineering Labs., Inc. Page 14 Manual No. M/94-310F III.2 RepeatabilityCAP 1000 and CAP 2000 Viscometers are repeatable to ±0.5% of the fu

Strany 7 - II. GETTING STARTED

Brookfield Engineering Labs., Inc. Page 15 Manual No. M/94-310Fd) If a printer is to be used, it should be connected (AC power & viscometer to pri

Strany 8 - CALIBRATE ? NO

Brookfield Engineering Labs., Inc. Page 16 Manual No. M/94-310F11. Read the results of the sample test on the printer or write down the test condition

Strany 9 - SAMPLE NO. 0000

Brookfield Engineering Labs., Inc. Page 17 Manual No. M/94-310F APPENDIX A - Cone Numbers - Sample Sizes - Viscosity RangesCAP 1000 AND 2000 VISCOMETE

Strany 10 - 100% = 2.500 P

Brookfield Engineering Labs., Inc. Page 18 Manual No. M/94-310FSAMPLE SIZEIt is necessary that sufficient sample is placed between the cone and plate

Strany 11 - II.8 Run and Stop Keys

Brookfield Engineering Labs., Inc. Page 19 Manual No. M/94-310FEXAMPLE: CAP 2000L Viscometer; Cone (02); running at 685 RPM;temperature 25.0°C; % FSR

Strany 12 - III. OPERATION

Brookfield Engineering Labs., Inc. Page 2 Manual No. M/94-310FTABLE OF CONTENTS I. INTRODUCTION ...

Strany 13

Brookfield Engineering Labs., Inc. Page 20 Manual No. M/94-310F APPENDIX B - Calibration ProceduresVerification of CalibrationNormally there is no req

Strany 14 - III.2 Repeatability

Brookfield Engineering Labs., Inc. Page 21 Manual No. M/94-310FFigure B1CALIBRATE ? NOToggling the INDEX key displays the words NO and YES.4. Pr

Strany 15

Brookfield Engineering Labs., Inc. Page 22 Manual No. M/94-310FPressing the CONE key after entering the fifth digit displays the message:Figure B4TEMP

Strany 16

Brookfield Engineering Labs., Inc. Page 23 Manual No. M/94-310FUpon pressing the CONE key to accept the last digit of the SMC value, the viscometer di

Strany 17 - CAP 1000 AND 2000 VISCOMETERS

Brookfield Engineering Labs., Inc. Page 24 Manual No. M/94-310FFigure B14CONE (00) CALIBRATIONVISCOSITY FSR TEMP S.STRESS S.RATE SPEED TIMER CONE SAMP

Strany 18

Brookfield Engineering Labs., Inc. Page 25 Manual No. M/94-310F APPENDIX C - Glossary of Screen Displays“CAL COMPLETE” - Cone calibration complete“C

Strany 19

Brookfield Engineering Labs., Inc. Page 26 Manual No. M/94-310F APPENDIX D - Variables in Viscosity MeasurementsAs with any instrument measurement, th

Strany 20

Brookfield Engineering Labs., Inc. Page 27 Manual No. M/94-310FViscometer Related VariablesMost fluid viscosities are found to be non-Newtonian. They

Strany 21

Brookfield Engineering Labs., Inc. Page 28 Manual No. M/94-310F APPENDIX E - CommunicationsPrinter Output - CAP 1000 & CAP 2000On the CAP Visc

Strany 22 - TEMP. = 25.0°C

Brookfield Engineering Labs., Inc. Page 29 Manual No. M/94-310FCAP 2000 TRANSMIT/RECEIVE COMMANDS FOR COMPUTER COMMUNICATIONNOTE: All multiplication

Strany 23

Brookfield Engineering Labs., Inc. Page 3 Manual No. M/94-310F I. INTRODUCTIONThe CAP Series Viscometers are Cone and Plate geometry high shear rate i

Strany 24 - CALIBRATING (00)

Brookfield Engineering Labs., Inc. Page 30 Manual No. M/94-310FLegend:<CR> - carriage return<ss> - status byte; returned as two hexadecima

Strany 25 - °C” - Main screen

Brookfield Engineering Labs., Inc. Page 31 Manual No. M/94-310FINTERPRETATION OF BYTE STATUS 7tiB 6tiB 5tiB 4tiB 3tiB 2tiB 1tiB 0tiBdeorezt

Strany 26

Brookfield Engineering Labs., Inc. Page 32 Manual No. M/94-310FSAMPLE PROGRAM FOR EXTERNAL COMPUTER CONTROL OF CAP 2000VISCOMETERThis program can be u

Strany 27

Brookfield Engineering Labs., Inc. Page 33 Manual No. M/94-310F1815 ‘ Send command to CAP 20001226 GOSUB 1870 ‘ Wait f

Strany 28

Brookfield Engineering Labs., Inc. Page 34 Manual No. M/94-310F1430 ‘1435 ‘1440 ‘ Routine used to insert a delay (approximately 50 milliseconds)1445 ‘

Strany 29

Brookfield Engineering Labs., Inc. Page 35 Manual No. M/94-310F1670 TINTERVAL = 301675 RETURN1795 ‘1800 ‘1805 ‘ Routine to send commands to the DV-III

Strany 30

Brookfield Engineering Labs., Inc. Page 36 Manual No. M/94-310F2000 ‘ If overflow, take the‘ complement2005

Strany 31 - Example:

Brookfield Engineering Labs., Inc. Page 37 Manual No. M/94-310F APPENDIX F - Warranty Repair and ServiceWarrantyBrookfield Viscometers are guaranteed

Strany 32

Brookfield Engineering Labs., Inc. Page 4 Manual No. M/94-310FThe following optional items may have been included:Part No.8. CAPCALC Software and PC C

Strany 33

Brookfield Engineering Labs., Inc. Page 5 Manual No. M/94-310FVISCOSITY TEMPERATURECONETIMERSTOPRPMINDEXRUNTEMPSAMPLEPRINTFigure 11) Set the viscomete

Strany 34

Brookfield Engineering Labs., Inc. Page 6 Manual No. M/94-310FSTOPThe STOP key has several functions:1. Stops the cone spindle rotation at any time.Th

Strany 35

Brookfield Engineering Labs., Inc. Page 7 Manual No. M/94-310F II. GETTING STARTED II.1 Power ONTurn the power ON using the switch located on the rear

Strany 36

Brookfield Engineering Labs., Inc. Page 8 Manual No. M/94-310FWhen using the solvent trap, connect it to the cone adapter by sliding it up, passing th

Strany 37

Brookfield Engineering Labs., Inc. Page 9 Manual No. M/94-310F II.3 Sample Number Setting (optional)Press the SAMPLE key and the screen will be displa