Brookeld Engineering Laboratories, Inc. Page 24 Manual No. M/07-022-D0613
TORQUE = 24.2%
MIN: 15 SEC:13
Figure II-25
NOTE: When this mode has begun, a press of the MOTOR ON/OFF key will
interrupt the time to torque operation and return the user to normal operation.
The seconds display will increment from zero (0) to 59 in one (1) second intervals and the
current value of the viscometer torque will be updated continuously. When seconds reach
59, the minutes value will increment by one (1) minute. This will continue until the user
selected torque value is attained at which point the viscometer will display the following
screen:
TORQUE = 85.0%
TIMED TORQ DONE
Figure II-26
At this point, the viscometer will stop the motor and continue to display this screen until
the user presses the
UP or DOWN ARROW keys to view the viscosity that was current at the
Timed Torque completion. The display would appear as follows:
22M 54S TO 85%
TIMED TORQ DONE
Figure II-27
CP 123456789
TIMED TORQ DONE
Figure II-28
The display will switch between that of Figures II-27 and II-28
for each press of either the
UP or DOWN ARROW key.
The test can be repeated by pressing the motor ON/OFF key.
Pressing any key except the UP or DOWN ARROW key or motor ON/OFF will cause the
viscometer to exit the Timed Torque mode and resume normal operation.
The time to torque value can be as high as 99 minutes and 59 seconds.