Brookeld Engineering Laboratories, Inc. Page 56 Manual No. M/07-022-D0613
MINERAL OIL VISCOSITY STANDARD FLUIDS
BEL Part No. Viscosity (cP) 25ºC
B29 29
B200 200
B600 600
B1060 1,060
B2000 2,000
B10200 10,200
B21000 21,000
B730000 73,000
B200000 200,000
B360000 360,000
Table E-2
Brookeld Viscosity Standard Fluid - General Information
We recommend that Brookeld Viscosity Standard Fluids be replaced on an annual basis, one year
from date of initial use. These uids are either pure silicone or mineral oil and are not subject
to change over time. However, exposure to outside contaminants through normal use requires
replacement on an annual basis. Contamination may occur by the introduction of solvent, standard
of different viscosity or other foreign material.
Viscosity Standard Fluids may be stored under normal laboratory conditions. Mineral oils should
be stored in the container in which they are supplied. Disposal should be in accordance with state,
local and federal regulations as specied on the material safety data sheet; MSDS information is
available upon request on our website to down load.
Brookeld Engineering Laboratories does not recertify Viscosity Standard Fluids. We will issue
duplicate copies of the Certicate of Calibration for any uid within two years of the purchase date.
Brookeld Viscosity Standard Fluids are reusable provided they are not contaminated. Normal
practice for usage in a 600 mL beaker is to return the material from the beaker back into the
bottle. When using smaller volumes in accessories such as Small Sample Adapter, UL Adapter,
Thermosel or Spiral Adapter, the uid is normally discarded.
Calibration Check Procedure for LV (#1-3) and RV,HA,HB (#1-6) Brookeld Spindles
NOTE: The LV #4 (#64) and the RV, HA, HB #7 spindle have been omitted from this
procedure. Brookeld does not recommend the use of these spindles to perform
a calibration check on your instrument. Reasons pertain to the small amount of
spindle surface area that makes contact with the viscosity standard, the difculty
of establishing the immersion mark precisely and the need for precise temperature
control at 25°C in the immediate vicinity of the spindle.