Brookeld Engineering Laboratories, Inc. Page 59 Manual No. M/07-022-D0613
Calibration Check Procedures for DIN Adapter
When a DIN UL Adapter is used, the water bath is stabilized at the proper temperature:
1) Put the proper amount of viscosity standard uid into the UL Tube. (Refer to the UL
Adapter instruction manual).
2) Attach the spindle (with extension link and coupling nut) onto the DV-I PRIME.
3) Attach the tube to the mounting channel.
4) Lower the tube into the water bath reservoir, or if using the ULA-40Y water jacket, connect
the inlet/outlets to the bath external circulating pump.
5) Allow sufcient time for the viscosity standard, sample chamber and spindle to reach test
temperature.
6) Measure the viscosity and record the viscometer reading; include % and cP (mPa•s).
NOTE: The spindle must rotate at least ve (5) times before readings are taken.
Calibration Check Procedure for a Helipath Stand and T-Bar Spindles
T-Bar spindles should not be used for verifying calibration of the DV-I PRIME Viscometer.
When a Helipath Stand and T-Bar spindles are used:
Remove the T-bar spindle and select a standard LV (#1-3) or RV,HA,HB (#1-6) spindle.
Follow the procedures in the beginning of this section, entitled Calibration Procedure for
LV (#1-3) and RV, HA, HB (#1-6) Brookeld spindles.
Calibration Check Procedure for Spiral Adapter
1) Place the viscosity standard uid (in the proper container) into the water bath.
2) Attach the spindle to the viscometer. Attach chamber (SA-1Y) and clamp to the viscometer.
3) Lower the DV-I PRIME into measurement position. Operate the viscometer at 50 or 60
RPM until the chamber is fully ooded.
4) The viscosity standard uid, together with the spindle, should be immersed in the bath.
Don’t introduce air bubbles.
Stirring the uid periodically (operate at 50 or 60 RPM), prior to taking measurements to
encourage temperature equilibrium.
NOTE: The spindle must rotate at least ve (5) times before readings are taken.
5) Measure viscosity and record the viscometer reading; include % and cP (mPa•s). Instrument
accuracy is ± 2% of the maximum viscosity range and not the standard 1%.