Brookfield DV2+Pro Viscometer Uživatelský manuál Strana 73

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 101
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 72
Brookeld Engineering Labs., Inc. Page 73 Manual No. M03-165-F0612
sample introduction techniques. The effect of these elements on measurement tolerance must be
consideredwhenverifyingthecalibrationofyourBrookeldviscometer.Sampletemperature
in all test circumstances is very important, and will also add an additional expanded tolerance
dependingonthetemperaturecontrolsystemandthecalibrationvericationtestsbeginwiththe
standard viscometer spindles as detailed above. Once the calibration of the viscometer itself is
conrmed,theexpandedtoleranceofthemeasurementsystemmaybedeterminedusingaccessory
devices. In many cases this additional tolerance will be very minimal, but as a general statement,
the addition of +/- 1% of the range in use is reasonable for accessories.