Brookeld Engineering Labs., Inc. Page 81 Manual No. M03-165-F0612
Calibration Procedure for a Small Sample Adapter
Brookeldrecommendsatwostepcheck.Firstverifythecalibrationoftheviscometerusingthe
standard viscometer spindles (LV #1-3, RV #2-6, HA #2-6 and HB #2-6 or cone/plate spindles) as
detailed in this appendix. Second verify the calibration of the viscometer using the Small Sample
Adapter. The use of an accessory device may increase the accuracy of measurement associated
with the DV-II+Pro.
When a Small Sample Adapter is used, the water jacket is connected to the water bath and the
waterisstabilizedatthepropertemperature:
1) Puttheproperamountofviscositystandarduidintothesamplechamber.Theamountvaries
with each spindle/chamber combination. (Refer to the Small Sample Adapter instruction
manual).
2) Place the sample chamber into the water jacket.
3) Putthespindleintothetestuidandattachtheextensionlink,couplingnutandfreehanging
spindle (or directly attach the solid shaft spindle) to the DV-II+Pro.
4) Allow 30 minutes for the viscosity standard, sample chamber and spindle to reach test temperature.
5) Measure the viscosity and record the Viscometer reading.
Note: The spindle must rotate at
least ve (5) times before readings are taken.
Calibration Procedure for a Thermosel System
Brookeldrecommendsatwostepcheck.Firstverifythecalibrationoftheviscometerusingthe
standard viscometer spindles (LV #1-3, RV #2-6, HA #2-6 and HB #2-6 or cone/plate spindles) as
detailed in this appendix. Second verify the calibration of the viscometer using the Thermosel.
The use of an accessory device may increase the accuracy of measurement associated with the
DV-II+Pro.
When a Thermosel System is used, the controller stabilizes the Thermo Container at the test
temperature.
1) Put theproper amountof HTviscosity standarduid intotheHT-2samplechamber.The
amount varies with the spindle used. (Refer to the Thermosel instruction manual).
2) Place the sample chamber into the Thermo Container.
3) Putthespindleintothetestuidandattachtheextensionlink,couplingnutandfreehanging
spindle (or directly attach the solid shaft spindle) to the DV-II+Pro.
4) Allow 30 minutes for the viscosity standard, sample chamber and spindle to reach test temperature.
5) Measure the viscosity and record the Viscometer reading.
Note: The spindle must rotate at
least ve (5) times before readings are taken.
Calibration Procedure using UL or DIN Adapters
Brookeldrecommendsatwostepcheck.Firstverifythecalibrationoftheviscometerusingthe
standard viscometer spindles (LV #1-3, RV #2-6, HA #2-6 and HB #2-6 or cone/plate spindles)
as detailed in this appendix. Second verify the calibration of the viscometer using the UL or DIN
Adapters. The use of an accessory device may increase the accuracy of measurement associated
with the DV-II+Pro.