Brookeld Engineering Labs., Inc. Page 95 Manual No. M03-165-F0612
Viscometer Will Not Return to Zero
Viscometer is not level
• Checkwithspindleoutofthesample
• Adjustthelaboratorystand
Pivot point or jewel bearing faulty
• PerformanOscillationCheck*
Remove the spindle and turn the motor OFF; select display to % torque mode.
Gently push up on the viscometer coupling.
Turn the coupling until the digital display reads 10-15 on the % display.
Gently let go of the coupling.
Watch the digital display; you should see a “run” of numbers next to the %; the “run” of
numbers should ultimately stop at 0.0 (+/- 0.1).
If the digital display does not return to ZERO, the unit most likely is in need of service.
• Performcalibrationcheck(SeeAppendixE)
• ContactBrookeldEngineeringLaboratories,Inc.oryourBrookelddealerforrepair(See
Appendix N).
* This procedure does not apply to instruments with ball bearing suspension. (See section I.3)
Display Reading Will Not Stabilize
Specialcharacteristicofsampleuid.Thereisnoproblemwiththeviscometer.
• RefertoAppendixC
Check for erratic spindle rotation.
• Verifypowersupply
• ContactBrookeldEngineeringLaboratories,Inc.oryourBrookelddealerforrepair.
Bent spindle or spindle coupling.
• ContactBrookeldEngineeringLaboratories,Inc.oryourBrookelddealerforrepair.
Temperatureuctuationinsampleuid.
• Usetemperaturebathforcontrol
No Recorder Response
Be Sure the viscometer is not at ZERO reading.
BesuretherecorderisONandnotonSTANDBY.
Verify the range settings.
Check cable leads for clean connection.
Verify cable connections (see Appendix H).