Brookfield DV2T Viscometer Uživatelský manuál Strana 30

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 95
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 29
Brookeld Engineering Labs., Inc. Page 30 Manual No. M13-167-B0614
QC Limits: Selectan acceptablerange for measurement results. The range may
be dened by Viscosity, Torque, Time, Temperature, or Shear Stress.
ThepossiblerangeforViscosityandShearStresswillbedenedbythe
spindle and speed selected. QC Limits are a visual and audible signal to
theoperatorduringthetest.Thedatasetdoesnotincludeanindication
of QC Limits violation.
AviolationoftheQCLimitsduringthetestwillbeindicatedbyaashingyellowboxaround
thedisplayforthespeciedparameter.NotethatTimedoesnothaveadedicateddisplay.
Aviolationwillalsoresultinanaudiblebeepandaonetimewarningmessage(seeFigure
III-6).Datawillcontinuetobecollectedwhilethewarningmessageisdisplayed.
Figure III-6
WhenViscosityorTorquearetheselectedQCLimitsparameter,QCLimitswillbeshown
duringthetestontheTrendBarifthatparameterisselectedfordisplay(seeFigureIII-7).
TheQCLimitsarerepresentedbythedashedlines.
TIP:AudiblealarmsmaybeturnedoffintheUserSettings
.
TIP: AnindicationofaQCLimitviolationisnotpartofthedataset.Theusercanrecord
aviolationbyusingtheNotesavailablewhenviewingtestResults(seeSection
III.9).
Trend Bar
Figure III-7
Zobrazit stránku 29
1 2 ... 25 26 27 28 29 30 31 32 33 34 35 ... 94 95

Komentáře k této Příručce

Sachin Vaijanath Ashture 05 Mar 2025 | 15:48:33

It's good information