Shear Rateiscalculatedfromtheselectedspeedbasedontheselectedspindle.
TIP:ShearRatewillbedisplayedaszeroforspindlesthatdonothaveSRCvalues.
TemperatureistheinputvaluefromaconnectedBrookeldtemperatureprobe.The
DV2TisprovidedwithaDVP-94Yprobethatcanbeinsertedintothetestsampleor
awaterbath.TheCone/PlateversionoftheDV2Tcanbeutilizedwithasamplecup
thatincludesanembeddedtemperatureprobe.SomeBrookeldaccessoriesinclude
temperatureprobes(Thermosel)oroptionalchamberswithembeddedtemperature
probes(SmallSampleAdapter).TheunitsoftemperaturearedenedintheGlobal
SettingssectionoftheSettingsMenu
.
Thetemperaturedisplaywillbe----whennotemperatureprobeisconnected.
SpeedistheselectedspeedfromtheCongureViscosityTestscreen.
TIP:SpeedcanbechangedbyswitchingtheRunningViscosityTestscreento
theViewTestscreenusingthecommandkeyatthebottomofthescreen.
TIP: When Speedis changed in the ViewTestscreen, theTest Namewill be
changedtounsaved.
Thestatusofthetestcongurationisindicatedinthegreybarabovethedataelds.A
congurationthathasbeensavedwillbeindicatedbyadisplayofthetestlename.A
congurationthatisnotsavedwillbeindicatedbyadisplayofUnsavedTest.
TheTrendIndicatorBarisdisplayedatthebottomoftheRunningViscosityTestscreen.
TheTrendIndicatorBarshowsthedirectionandmagnitudeofchangeinthemeasured
parameter.ThisbarcanrepresenteitherthemeasuredViscosityorthemeasuredTorque.
Tochangethedisplayedparameter,touchthedataeld.InFigureIII-7,Torqueisshown
intheTrendIndicatorBar.Thisisrepresentedbyabluecirclebesidethetorquevaluein
thedataeld.WhenQCLimitsareselectedtobeeitherTorqueorViscosity,thelimitswill
beindicatedontheTrendIndicatorBar(seeSectionIII.7).
TwoCommandKeysareavailableontheRunningViscosityTestscreen:ViewTestand
StopTest.
Stop Test:Immediatelystopsthecurrenttest.Theuserwillbepresentedwitha
conrmationbox.ThetestwillcontinuetorununtilYesisselectedinthe
Conrmationbox.Ifanydatahadbeencollectedduringthetest,theuser
willbepresentedwiththeResultsscreen(seeSectionIII.9).Ifdatawasnot
collected,theuserwillbereturnedtotheCongureViscosityTestscreen.
TIP:Iftheoperatorselects“Donotshowthismessageagain”withinthe
Conrmationbox,thentheConrmationboxwillnolongerbedisplayed
whenStopTestispressed.ThisconditioncanberesetintheDisplay
section of User Settings
.
View Test:Change the view from measurement to test parameters including:
Spindle,Speed,Temperature,EndCondition,DataCollection,QCLimits,
Instructions,andDensity.TheViewTestscreenincludesalivedisplay
ofTorqueandaTimeparameterthatshowsthetotalelapsedtimeforthe
test(seeFigureIII-10).